TOC Analyzer | Benefits for selecting the PAT700

 
TOC PAT700
 

The PAT700 TOC analyzer fully complies with all global pharmacopeias (USP, EP, JP) requirements. Each sample is fully oxidized guaranteeing the most accurate TOC result that follows EP2.2.44 guidelines. Production uptime is assured with the PAT700 main and standby dual UV lamp design, making it the ideal choice for on-line TOC release. Save time and money - the PAT700 can be installed online and run up to 4 grab samples offline.

 

 

 

 

 

 

PAT700 TOC Analyzer Product Specifications

TOC Operating Range 0.5 to 2,000 ppb as Carbon
  Display Resolution 0.1 ppd
  Accuracy +/-1 ppd or +/-5% whichever is greater
  Repeatability +/-0.3 ppd or +/-1% whichever is greater
  Limit of Detection 0.5 ppb
  Maximum Input Conductivity 0.2 uS/cm for all waters, 1.0 uS/cm for all neutral waters, 5.0 uS/cm for water with CO2 as the sole conductive species
Conductivity Conductivity Range 0.05 to 150 uS/cm (@25* C)
  Display Resolution 0.01 uS/cm
  Conductivity Accuracy +/-2% over full range (uncompensated)
  Resistivity Resistivity Range 0.2 to 18 MQ-cm (@25*C)
  Display Resolution 0.01 over full range
  Available Modes Temperature compensated to 25*C, or uncompensated
Temperature Ambient Operating Range  10 to 40*C (50 to 104*F)
  Measurement Accuracy +/-0.4*C
  Sample Water Range 5 to 95*C (41 to 203*F)
  Display Resolution 0.1 over full range
Physical Specs UV Lamps 2, with UV Detect technology
  Interface/Display Color with touch screen
  Maximum Altitude 4,000 m (13,125)
  User I/O Wiring Three, 3/4-inch conduit openings or quick disconnect fittings
  Standards System Onboard, Automated Standards Introduction System (OASIS)
  Dimensions 59.7 w X 22.9 d X 25.4 h (23.5 X 9 X 10 inches)
  Weight 13.6 kg (30 lbs)
  Sample Inlet Flow Rate Range 60 mL/min to 300 mL/min
  Sample Inlet Pressure Range 10 to 100 psi (69 to 690 kPa)
Compliance Installation Category II
  Pollution Degree 2, IEC 61010-1
  CE Compliance EN 61010-1 and EN 61326
  Safety Rating ETL, conforming to UL 61010-1 and CSA 22.2 No. 61010-1
  Enclosure Rating Conduit version: IP56 Quick connect version: IP46
  Release Tests USP <643>, USP <645>, JP 16, EP 2.2.44
New Features CIP Selectable mode for Clean-In-Place analysis
  Dual Stream option Toggle or programmable stream switching
  Excursion sampling Minimum flow rate to fill excursion bottle = 160 mL/min
  Rouge detection Identifies oxidation cell contamination from roughing

ASX 7200 Auto Sampler  Specifications

Physical Specs Vial Capacity 36 Grab sample bottles HDPE or Borossilicate USP Type 1 glass
  Dimensions 45.7 cm X 36.6 cm X 53.7 cm (HxWxD)
  Weight 21 kg (45 lbs)
Communications Communications RS232 interface with PAT700 - No external PC required
Environment Ambient Environment 10 - 40*C, 5 - 95% non-condensing RH
Power Power Requirements 100 - 240 VAC +/- 10%, 4.4A


สนใจสินค้าและบริการต่าง ๆ กรุณาติดต่อ

บริษัท แอดวานซ์ เมทโทรโลยี จำกัด 

Tel: 089-8233773 # 083-7790808 # 086-3124690 # 02-0153960   Fax : 02-0153961

LINE OA : @adm-adm

E mail : a@adm-adm.com 

E mail : adm@adm-adm.com

E mail : admin@adm-adm.com

E mail : info@adm-adm.com 

E mail : md@adm-adm.com 

E mail : nui_cal@hotmail.com 

www.adm-adm.com

www.adm-advance.com

0 บาท
จำนวน:
Visitors: 711,557